FLOWMATTERS™ NEWSLETTERS
Get the latest information on a variety of applications, case reports, studies, and trials. Subscribe Here >>

 

 

Mean Flow

Assess Mean Flow to Confirm Graft Patency

Normal 30 ml/min (or higher) Mean Flow =  Patent Graft

See Table, p. 21 of CABG handbook for a compilation from the scientific literature of the “Normal Flow Range” for various coronary grafts.

Flow under 5 ml/min  = Graft in Trouble

Look for kinks, twists in the graft, low MAP, vasospasm.

Redo anastomosis if indications point to technical error.

 

Factors Affecting Flow


Look for kinks, twists in the graft, low MAP, vasospasm.

Redo anastomosis if indications point to technical error.

 

Assess other factors that may account for a lowered flow:

 

Further Reading