Transonic Systems Inc.
top_right

Publications | Site Map

 
 
Research Surgery Critical Care Hemodialysis Radiology OEM About Contact FAQ's
 

 

 

 

 

 

PXL-Sterile Tubing Flowsensors

For Extracorporeal Applications

 

    • XLArtificial Heart & VAD Performance
    • Medical Device & Pump Engineering
    • Manufacturing & Compliance Flow Testing

Unique ultrasonic transit time technology measures volume flow with sensors specifically designed for greatest accuracy on tubing.

Specifications
Acoustics
Calibration
Ordering

Most non-aerated liquids can be measured including saline and buffer solutions, blood, water and even Diesel fuel. No physical contact is made with the fluid media. PXL-Series Flowsensors can be calibrated and programmed for up to 4 different fluid / temperature / tubing combinations and will work with most flexible tubing types. Sensor size is determined by outside diameter of the tubing.

SENSOR SIZE
ME-

APPLICATIONS

FLUID TYPE

-2PXL

Infusion pumps, intravenous tubing, perfusion devices, delivery systems

cell culture, saline, buffer

-3PXL

Bioreactor tissue generation, isolated organs & vessels, perfusion devices/delivery systems, shunts

cell culture, saline, buffer

-4PXL

Experimental dialysis, bioreactor tissue generation, perfusion devices/delivery systems, isolated organs/vessels, shunts

cell culture, saline, buffer

-5PXL

Experimental dialysis, bioreactor tissue generation, perfusion devices/delivery systems, isolated organs/vessels, shunts

cell culture, saline, buffer

-6PXL

Experimental CP Bypass, ECMO, organ perfusion

blood, water/glycerine
saline

- 7PXL

Experimental CP Bypass, ECMO

blood, water/glycerine
saline

-8PXL

Experimental CP Bypass, ECMO

blood, water/glycerine
saline

-9PXL

Experimental CP Bypass, ECMO

blood, water/glycerine
saline

-10PXL

VAD (Ventricular Assist Device)

blood,
blood analog, water/glycerine

-11PXL

VAD (Ventricular Assist Device)

blood analog, water/glycerine

-12PXL

VAD (Ventricular Assist Device)

blood analog, water/glycerine

-14PXL

VAD (Ventricular Assist Device)

blood analog, water/glycerine

-16PXL

TAH (Total Artificial Heart), VAD (Ventricular Assist Device

blood analog, water/glycerine

-20PXL

TAH (Total Artificial Heart), Mock Circulatory Systems, Heart Valve Tubing

blood analog, water/glycerine

*Note: These are typical flow rates requested and are not indicative of the maximum or minimum flow rate. Fluid types listed are typically requested and not exclusive to size.

 

Calibrated to Specification

Sterile tubing flowsensors are calibrated for the tubing fluid, temperature and typical flow rate with which the sensor will be used for the greatest accuracy. Up to four combinations are possible for each sensor.
 

Fluid Type & Temperature

Since fluid density and temperature alter the transit time of the ultrasound signal, the sensors are calibrated for these parameters. Sttandard calibrations include: Blood, glycerine/water solutions, saline and buffer solutions at 23 or 37 degrees C.Custom calibration for other solutions is available for extra charge. (An MSDS is required before handling the solution).
 

Tubing Type

Medical grade and laboratory tubing are generally compatable for use with XL-Sensors. Rigid PVC and lucite are not. We stock several standard tubing types noted in the table below. For other tubing types, a 30 cm sample is required for sensor calibration.
 

Calibration Certification

Transonic Systems provides documentation for calibration traceable to NIST standards.
 

Stocked Tubings

Probe Size
ME-

Tubing (inches)

Stock Tubing

OD

ID

Wall

-2PXL

1/8

3/32

1/32

Tygon R-3603

-3PXL

3/16

1/8

1/32

Tygon R-3603

-4PXL

1/4

 

 

No stock tubing

-5PXL

5/16

3/16

1/16

No stock tubing

-6PXL

3/8

1/4

1/16

MEDIFLEX, Tygon R-3603,

-7PXL

7/16

1/4

3/32

MEDIFLEX, Tygon R-3603

-8PXL

1/2

3/8

1/16

Tygon R-3603

-9PXL

9/16

3/8

3/32

BENTLY BYPASS, MEDIFLEX,
Tygon R-3603

-10PXL

5/8

1/2

1/16

Tygon R-3603

-11PXL

11/16

1/2

3/32

BENTLY BYPASS 70, Tygon R-3603

-12PXL

3/4

1/2

1/8

Tygon R-3603

-14PXL

7/8

5/8

1/8

Tygon R-3603

-16PXL

1

3/4

1/8

Tygon R-3603

-20PXL

1 1/4

1

1/8

Tygon R-3603

 

 

 
Start receiving our
Flow Matters™ Newsletter.









 

 

 

 

 

 

Blue_Printer90

Print Page

 

 

 

 

 

 

 Surgery | Critical Care | Research | Hemodialysis | OEM | Publications | Customer Service

 

 