|
HD02 Troubleshooting Case
A nurse from a dialysis clinic called with questions about some HD02 measurements on a particular patient. She explained the
situation as this:
In Normal line Position: Dialysis Machine Pump Flow = 400 ml/min, Delivered Flow by HD02 = 160 ml/min and Access
Recirculation = 0%.
In Reversed Line Position: Dialysis Machine Pump Flow = 300 ml/min, Delivered Flow by HD02 = 160 ml/min and Two Access
Flow Measurements = 1800 ml/min and 2000 ml/min.
Question 1: What was the delivered flow reading on previous patient? Qb by HD02 and HD machine corresponds
within 10%.
Question 2: What was the access flow reading on this patient the previous month? 2000 ml/min at Qb = 300ml/min.
Possible Solution: Check blood line kinking. Response: Yes, bloodline was kinked near dialyzer.
Message: Check bloodline tubing for occlusion when there is large discrepancy between pump flow and delivered flow.
Believe in the HD02 measurements!
Important Note on Yearly Calibration of HD Systems
It is important to send HD systems in for a yearly calibration. This will certify the accuracy of the system and allow
reproducibility of the measurements. A
calibration reminder is built into the HD02 software. Under MDD law, calibration must be done by someone certified to do this. Transonic is the only firm which can perform such
certifications.
HD02 Software Tip
When there is “No Recent Measurement Detected”
Whenever the traffic light turns green, the software will detect any idle use of the program for 10 minutes. During the last
minute, an on-screen countdown will start. The options are Continue with Current Patient or Stop Patient Session. If neither option is selected, the patient session will stop
automatically after the 60 second countdown. Since the HD02 software is always recording waveforms to disk when in a patient session, this feature is a precaution to prevent filling up
the hard disk with essentially no data. It also prevents the user from accidentally testing another patient in the same session.
|