FlowQCHT110 Bypass Flowmeter

An Independent Measure of Volume Flow through Medical Tubing

 

 

HT110
HT110 Bypass Flowmeter & HXLSensor

Designed for ease of use in the critical care setting, the HT110 Bypass Flowmeter and Sterile Tubing Flowsensor monitor volume flow through medical tubing in intensive care and outside the surgical field in the operating room. Clamp-on sterile tubing sensors do not compromise tubing sterility or non-thrombogenic surfaces.

By providing reliable, immediate volume flow measurements, the Bypass Flowmeter system enhances safety for the patient by insuring accurate delivery in procedures using roller pumps, centrifugal pumps and pressuredriven circuits.

Flowsensors

Transonic XL Sensors
Ultrasonic transit-time sensors, specifically designed for tubing, measure volume flow with highest accuracy. Most non-aerated liquids can be measured including saline and buffer solutions, blood, water and even diesel fuel. No physical contact is made with the fluid media.

XL-Series Flowsensors can be calibrated and programmed for up to 4 different fluid / temperature / tubing combinations and will work with most flexible tubing types. Sensor size is determined by outside diameter of the tubing.

 

Support

Operator’s Manual

Service Manual
 

Specifications

Weight/Size HT110: single channel 7 lbs. (3.2 Kg), 8 1/2" w x 3 1/2" h x 13" d
HT110: 90-265 Volts; 50-60Hz, single phase, 40 Watts, (1.5 Amp fuse)
Electrical Isolation: Flowmeter complies with USA standards for medical and dental equipment (UL544), and with European standards for medical and ultrasonic apparatus
(DIN IEC 601-1, VDE 0750 -1/5.82, IEC 62D Sec. 31).
Alpha-Numeric Display: Single row by 16 character liquid crystal that displays error conditions, calibration and probe parameters.
Flowmeter Display: 0.1 ml/min to 200l/min (typical), displayed by a 4 1/2 digit, 7 segment LED.
Flow Monitor Outputs: Analog mean and pulsatile flow
Probe Connector: 15 pin, high density D-Sub connector
Digital Identification: Probe identification & calibration parameters programmed into flowprobe connector
Ultrasonic Frequency: 300 KHz to 7.2 MHz
Accepts: XL-Series sterile tubing flowsensors for extracorporeal tubing
 

Electrical Specifications

Power AC Input: 90 - 265 VAC; 50 - 60 Hz, 40 watts; Fuses: Two, 1.5A / 250V, 5 x 20 mm
Connector: International 3 conductor type IEC 320
Flow Monitor Average: Grounded BNC fitting ± 5.0 volts peak, Output impedance 500 ¥; Filtered at .1Hz. Pulsatile: Grounded BNC fitting ± 5.0 volts peak, Output impedance 500 ¥; Filtered at 5Hz
Bubble Switch Isolated ground BNC fitting connected to SPST solid state relay, optically coupled, normally closed

Voltage Isolation: 1000 Vdc
Max switch voltage: 40 Vdc
Max switch current: 0.5 Amps continuous
Max switch power: 0.35 watts
Typ. ON resistance: 1 ohm
RS-232: 9-pin D-Sub connector, standard RS232 parameters
Isolated ground: pin 5; Transmit: pin 2; Received: pin 3
 

Computer Option

The (-P) Personal Computer Interface Option allows the user to interface the flowmeter to a standard IBM-PC computer to display, record and playback flow waveforms and data.
Package includes: Additional circuitry within the flowmeter cabinet;

  • a standard RS-232 cable
  • WinDaq - PC compatible software to display &
    record flow data.

 

 

 