Transonic Systems Inc.
top_right

Publications | Site Map

 
 
Research Surgery Critical Care Hemodialysis Radiology Bioprocess OEM About Contact FAQ's
 

 

 

 

 

 

ANNA Presentation Abstracts

 

[SU-PO363] Cardiac Output (CO) Monitoring during Haemodialysis (HD) a Comparison between Saline Dilution Method (Transonic), and a Noninvasive Method Based on Rebreathing of Inert Gas (Innocor).

Danny Jensen, Jens Dam Jensen. Dept. of Nephrology C, Skejby, Aarhus University Hospital, Aarhus, Denmark

Monitoring of CO during haemodialysis (HD) is of great interest, due to the cardiovascular effects of HD. Innocor (I) is a new, non invasive portable device for assessing CO with built in pulse oxymeter and multigas analyser. The device does all analyses of gas automatically on site. The gasses used in the I device are: Sulfurhexaflouride (SF6) 0,1% (measures lung volume) and N2O 0,5 % (measures pulmonal blood flow) and O2 25% (variable). We compared CO measurement by I with Transonic (T) in a controlled trial in which the sequence (T or I) was randomised. Twentyfive (m/f = 18/7) stable chronic HD pts age mean = 54(24 79) y were investigated. Dry weight = 68(42 - 96) kg. Mean patient access flow was 1110 134(360 2775) ml/min. Dialyser blood flow rate = 289 10 (192 404) ml/min. Mean Hct = 0,38 (0,31-0,46). Mean O2 uptake = 0,21 0,087 (0,08-0,49) L/min. Pulmonal shunt fraction = 8 8,4 (0 41) %. All patients had stable A-V arm fistula with no local fistula recirculation. We found a CO T = 5,9 0,35(3,9 - 10,5) L/min. CO I = 4,9 0,34(1,8 - 9,2) L/min. Mean delta CO (T I) = 1,0 0,2(1,4 - 0,6) L/min (p< 0.001). The measurement repeatability of both methods showed an identical coefficient of variance (sd/mean CO), CVT = 6%, CVI = 7% in both high and low CO levels. Pearsons linear correlation coefficient (COT vs. COI) = 0,84 (p<0.001). A linear regression showed COI = 0,99*COT1 (r2 = 0,7). Nine (36 %) of our observations on delta CO were within 15% limits of agreement in a Bland - Altman plot.
Access flow in % of CO correlated significantly with delta CO, Spearman rho = 0,68 (p<0,01).
Conclusion: We found a clear relationship between CO measured with T and I. Both demonstrated reproducible results. The I underestimates CO compared with T. This may be due to rapid recirculation of N2O through the A-V fistula with interference of pulmonary test gas exchange. The association between the difference in CO measured by T vs. I - and access flow in % of CO supports this hypothesis.

Sunday, October 31, 2004 , 10:00 AM

Poster: Cardiovascular Disease in Dialysis: Clinical Aspects (10:00 AM - 12:00 PM) Poster Board Number: SU-PO363


[SU-PO368] Blood Flow in Brachiocephalic AVF Represents a Greater Percentage of Cardiac Output Than That in Radiocephalic AVF.

Wama van der Mark, P. Boer, P.J. Blankestijn. Department of Nephrology, University Medical Center Utrecht, Utrecht, Netherlands

The flow in the arteriovenous (AV) access of hemodialysis patients means an extra burden to the cardiac output (CO) and is therefore a risk factor for left ventricular hypertrophy and heart failure. The aims of this study were to quantify access flow and to assess the percentage of CO that passes through the access and to compare the different types of accesses in this respect.
During a single hemodialysis session both CO and access flow were quantified by the Transonic indicator dilution technique.
We performed 78 sets of measurements; 35 in AV fistulae forearm (radiocephalic), 29 in AV fistulae upper arm (brachiocephalic), and 14 in forearm PTFE grafts (in all cases from brachial artery to cephalic vein).
In forearm AV fistulae, upper arm AV fistulae and PTFE grafts, CO (meanse) was 5351315 mL/min, 6541364 mL/min and 4661349 mL/min (p<0.01 upper arm AV fistulae versus others). Access flow was 966114 mL/min, 1693143 mL/min and 71577 mL/min (p<0.001 upper arm AV fistulae versus others). Access flow represented 16.91.3%, 26.72.1% and 15.51.4 % of CO. The percentage in upper arm AV fistulae was significant higher (p<0.05) than in the two other groups. Blood pressure during measurements did not differ between groups, indicating that vascular resistance in the arms of patients with upper arm AV fistulae was lower than that in the other two groups.
In conclusion, both in absolute terms and relative to CO flow in upper arm AV fistulae is higher than in forearm AV fistulae and PTFE grafts. This is a result of the lower vascular resistance in the arms of these patients. High CO may mean an additional cardiovascular risk factor. The results underscore the preference for radiocephalic AV fistulae over brachiocephalic AV fistulae as primary choice for vascular access.

Sunday, October 31, 2004 , 10:00 AM

Poster: Cardiovascular Disease in Dialysis: Clinical Aspects (10:00 AM - 12:00 PM) Poster Board Number: SU-PO368


 

[SA-PO309] High Flow Fistula and Cardiac Hemodynamics.

Jennifer M. MacRae, Thuan H. Do, Debbie Rosenbaum, Adeera Levin, Mercedeh Kiaii. Nephrology and Internal Medicine, University of British Columbia, Vancouver, BC, Canada

AVF creation is associated with a decrease in peripheral resistance and increased cardiac output (CO). High output cardiac failure (HOCF), a rare complication of high flow AVF, is defined as symptoms of CHF in the presence of a high cardiac index ( 3.0 L/min/m2). This study was to determine the prevalence of high flow AVF, to document pt and AVF characteristics associated with high flow, and to compare hemodynamic and echo parameters of high and normal AVF flow pts.
High access flow, Qa, was defined as > 2.0 L/min 10% and normal Qa as 600 1500 ml/min. Qa, CO, central blood volume (CBV) and total peripheral resistance (TPR) were measured in duplicate using ultrasound dilution technique (HD02 Monitor Transonic Inc.) on the mid week dialysis run in chronic stable HD pts. Echo was performed on the same day .
19 % (46/242) of the AVF population had Qa> 2.0L/min. We report the results on 21 patients with high Qa and 15 with normal Qa. High Qa is associated with a younger age (55y vs 73y, p<0.05), upper arm AVF (p=0.059), hx of a previous AVF(p<0.05) and a higher PTH level. Table 1 shows the hemodynamic and echo data for the two groups.
Both CO and Qa/CO ratios are significantly higher in the high Qa pts. In these pts we do not know the overall risk of CV outcomes or at which degree of Qa or Qa/CO this may occur. Prospective monitoring of Qa/CO in high flow pts is necessary to better understand their potential cardiac risk and to develop intervention strategies.

 

BP mmHg

CO* L/min

TPR*

Qa* ml/min

Qa/CO*

LVMI

LVEF %

Normal Flow AVF (15 pts)

135/70

5.0(4.4-5.7)

17.8(16-20)

925(660-1110)

16(13-25)

102(93-142)

65(65-70)

High Flow AVF (21 pts)

143/76

8.1(6.5-14.8)

12.5(11-15.3)

2260(2100-3750)

32(26-38)

122(114-149)

65(60-65)

*p<0.05


Funding Source: Kidney Foundation of Canada

Saturday, October 30, 2004 , 10:00 AM

Poster: Managing Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: SA-PO309

 

 


[PUB239] Variation of Hemodynamic Parameters during Hemodialysis Assessed by the Transonic Flow Method.

Jens-Uwe Kriete, Bernd Sucke, Werner Riegel. Medizinische Klinik III, Klinikum Darmstadt, Darmstadt, Germany

Cardiovascular diseases are the leading cause of mortality in end stage renal disease patients. Hypotensive episodes during hemodialysis have a major adverse impact on the survival rate of hemodialysis patients. The main hemodynamic compensatory mechanisms in patients undergoing dialysis are largely unknown. The aim of this study is to investigate the intradialytic stability of dialysis patients by measuring hemodynamic parameters. Cardiac Output (CO), Cardiac Index (CI) and Peripheral Resistance (PR) are assessed by means of Transonic Hemodialysis Monitor HD02.
We monitored CO, CI and PR every hour during dialysis sessions using Transonic Hemodialysis Monitor HD02 (Diatec GmbH, Gilching, Germany). Eleven stable patients who had been on renal replacement therapy for more than one year were investigated. The achievement of dry body weight after dialysis was assessed by ultrasound examination of the vena cava.
Data are given as mean SD. Nine patients showed a decrease in CO (1.38, 0.75 l/min) and CI (0.73, 0.35) and a parallel increase in the PR (6.56, 3.1 mmHg/l/min). The ultrafiltration rate of each patient varied between 362.5 ml/h and 1300 ml/h. The ultrasound investigation of the vena cava after dialysis showed inspiratory collapse in these 9 cases. We did not detect blood recirculation in the vascular accesses of any patient. We noticed an increase in the CO (1.35 l/min), CI (0.85) and a decrease of the PR (4.55 mmHg/l/min) in 2 patients. A collapse of the vena cava was not observed in these cases.
The Transonic Monitor HD02 is a convenient device for monitoring cardiovascular parameters during dialysis. Patients who attained their dry body weight after dialysis showed a moderate decrease in cardiac output and cardiac index and a related increase in the peripheral resistance. These findings were not correlated to the ultrafiltration volume. The observed increase in cardiac output and cardiac index in patients who do not reach their dry body weight could be interpreted as an improvement of heart function due to recompensation of fluid homeostasis.
Disclosure - Grant/Research Support: Fresenius Medical Care AG, Bad Homburg, Germany



Publication Only


[F-PO403] Monitoring of Dialysis Access by Transonic Flow Studies Is Operator Dependent.

Jeffrey I. Silberzweig, Kathryn Osborn, Yvonne Noak. Hemodialysis, The Rogosin Institute, New York, NY; The Division of Nephrology, Weill Medical College of Cornell University, New York, NY

Reductions in flow lead to subtle often unrecognized compromises in delivery of dialysis and presage failure of the access device itself. Access flow can therefore be considered a fundamental property of the access that should be monitored1. K/DOQI guidelines call for surveillance of AV grafts by intra-access flow and referral for fistulogram if the flow is less than 600 mL/min or there is a decrease in flow by more than 25% over 4 months2.
We initiated an intra-access flow monitoring program using the Transonic monitor in May 2002. Our program aimed to increase the sensitivity of the test by using a cutpoint value of a 10% decline in flow over one month. Between January 2002 and April 2004 (excluding January 2003 and February 2004), 1508 Transonic studied were done. Of these, 239 showed at least 10% decline in flow from readings obtained one month earlier. Of these, 127 were referred for angiograms; 127 (100%) patients had abnormal stenotic abnormalities requiring percutaneous transluminal angioplasty (PTA). Among the 1269 patients with normal studies, 4 (0.32%) had a graft thrombosis. These data indicate a sensitivity of 96.95% and a specificity of 100%.
The success of our monitoring program led to a dramatic reduction in dialysis graft thromboses. Patients profited by a marked reduction in the need for dialysis access via femoral catheterization. During the six months prior to initiating the monitoring program, 17 patients required femoral catheterization for access thromboses; two years later, only 2 such procedures were done.
References:
1. Depner, TA, et.al., ASAIOJ, 541, 1995, 745.
2. National Kidney Foundation, AJKD, 37 Suppl., 2001, S150

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO403


[SU-PO270] Cardiovascular Stability of Intermittent Single Pass Batch Dialysis a Prospective Study Using the Transonic Flow Method.

Bernd Sucke, Jens-Uwe Kriete, Werner Riegel. Medizinische Klinik III, Klinikum Darmstadt, Darmstadt, Germany

Single Pass Batch Dialysis (Genius) is a successfully established renal replacement technique in end stage renal disease patients. It is supposed to be associated with less severe decline in blood pressure and lower increase of pulse rate compared to conventional hemodialysis.
This study compares the intradialytic hemodynamic changes in patients undergoing Genius dialysis and conventional intermittent hemodialysis.
Ten patients (3 women, 7 men) received alternating intermittent Genius dialysis and conventional hemodialysis (Fresenius 4008, each Bad Homburg, Germany) randomly. Patients had been on chronic renal replacement therapy for more than one year. Cardiac Output (CO), Cardiac Index (CI) and Peripheral Resistance (PR) were monitored once an hour during the dialysis session. Measurements were performed by the Transonic Flow method (Diatec GmbH, Gilching, Germany) using a hemodilution technique. Prior to dialysis a recirculation in the vascular access was excluded. The achievement of dry body weight after dialysis was controlled by ultrasound investigation of the inferior vena cava.
Data are given as mean SD. Eight patients showed a decrease in CO (F4008 vs Genius: 1.21 0.59 vs1.3 0.69 l/min), CI (0.66 0.3 vs 0.71 0.39) and an increase in PR (5,77 2.28 vs 5,53 3.95 mmHg/l/min) at the end of the dialysis session. These 8 patients achieved their dry body weight at the end of dialysis. Two patients showed an increase in CI and a decrease in PR during conventional dialysis. Both patients failed to attain their dry body weight. Remarkably, 4 patients showed a transitory decline in PR (2 1.08 mmHg/l/min) after starting Genius dialysis session.
Focussing on cardiovascular stability, intermittent Genius dialysis seems to be an appropriate alternative to conventional dialysis. There are no significant differences of the hemodynamic parameters during dialysis comparing both techniques of renal replacement therapy when dry body weight at the end of dialysis was achieved. The transitory decline of the PR in 4 patients receiving Genius dialysis may be explained by the contact of blood to warm dialysate at the beginning of the treatment.
Disclosure - Grant/Research Support: Fresenius Medical Care AG, Bad Homburg, Germany

Sunday, October 31, 2004 , 10:00 AM

Poster: Hemodialysis: Body Composition and Hemodynamics (10:00 AM - 12:00 PM) Poster Board Number: SU-PO270


 

[F-PO398] Myths and Realities of Vascular Access Surveillance Programs during Hemodialysis (HD).

Nikolai M. Krivitski. Engineering, Transonic Systems Inc., Ithaca, NY

Since the introduction of intra-access pressure and flow measurement techniques, a large body of literature on access surveillance has been produced. Recently, the efficacy of access surveillance has been debated. The purpose of this abstract is to summarize papers on access surveillance techniques with respect to the ability to identify stenosis and thrombosis. Criteria for inclusion: 1) papers from peer-reviewed journals or books (no abstracts); 2) independent studies only--papers by authors and companies that invented the surveillance technology were not included. Results are presented in the Table.

Summary of results of vascular access surveillance trials

Surveillance

Dynamic venous pressure

Static venous pressure

Access flow during HD

Independent studies

10

7

22

Number of accesses

AVG=748 AVF=150

AVG=353 AVF=84

AVG=1501 AVF=1246

Investigated stenosis

positive=4* negative=1*

positive=3* negative=0*

positive=11* negative=0*

Investigated thrombosis

positive=4* negative=6*

positive=2* negative=4*

positive=12* negative=5*

Major problems with some studies and technologies

Wrong basic theoretical assumption

Wrong basic theoretical assumption

Not following both K/DOQI threshold guidelines; PTA failure

* authors conclusions on surveillance outcomes; only papers included-no abstracts; some studies had both positive and negative conclusions on stenosis and/or thrombosis.

Among 10 studies that used only one flow threshold, a positive conclusion was reached in 7 studies and a negative conclusion was reached in 3 studies. Among 12 studies that used both absolute and trended access flow thresholds (analogous to the K/DOQI guidelines), 10 had positive conclusions about flow surveillance and stenosis detection or thrombosis prediction. Two presented negative conclusions--one identified PTA failure, and the other had results that are in dispute.
The summary of 17 independent intra-access pressure clinical studies does not support the statement that static pressure is a better surveillance tool than dynamic venous pressure. Analysis of 22 access flow studies shows that it is more efficacious to use both absolute and trended thresholds, as promoted by the K/DOQI guidelines, to achieve positive outcomes.
Scientific Advisor: Employee of Transonic Systems Inc.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO398

 

 


[SU-PO275] Hemodynamic Stability during Pre-HDF and Lowflux HD at Temperature Controlled Conditions Using High Calcium-Ion Dialysis and Replacement Fluid. A Blinded Randomized Controlled Study.

Danny Jensen, Nikolaos Karamperis, Jens D. Jensen. Dep. of Nephrology C, Skejby, Aarhus University Hospital, Aarhus, Denmark

Convective therapy is thought to result in better hemodynamic stability than conventional low flux hemodialysis (HD). We compared HD with high dose pre-hemodiafiltration (HDF) in 13 dialysis patients (pts), mean age 49 y (26-76). Pts with NYHA > II were excluded. Pts were randomized to either HD followed by pre-HDF (4 h) or vice versa. The dialysis modality was blinded. Substitution fluid volume in HDF was 1,2 L/kg dry weight (83 2L). Ca-ion fluid concentration was 1,75 mM. Arterial temperature (atmp) was locked at pt temp. at dialysis start. Cardiac output (CO) was measured hourly during the sessions using the Transonic Flow QC saline dilution method.
Mean blood pressure (MBP) change during HD (end vs. start) was -0,52 2,6 mmHg (ns) and -2,8 3,1 mmHg (ns) during HDF. MAP decrease HD vs. HDF (ns). CO decreased during HD from 6,9 0,5 L/min to 6,1 0,44 L/min (p<0,05) and during HDF from 7,4 0,24 L/min to 5,9 0,52 L/min (p<0,01). CO decrease in HD vs. HDF (ns). The stroke volume (SV) during HD declined 5 6 mL (ns) and 14 6 mL (p<0.05) in HDF. SV decline HD vs. HDF was 10 6 mL (ns). Total periferal resistance (TPR) increased in HD by 2,2 1,2 mmHgmin/L (ns) and in HDF by 2,8 0,7 mmHgmin/L (p<0,01). The differences in TPR increase in HD vs HDF was 0,6 1,1 mmHgmin/L (ns). Pulse rate (P) during HD decreased 6 2 bpm (p<0,05) and 3 5 bpm (ns) during HDF. Diff. in P decreases between modalities (ns). Energy loss (cooling) during HD = 228 41 kJ and during HDF = 207 40 kJ (ns). The atmp was kept constant 36,6 0,1C in HD and 36,5 0,1C in HDF. Ultra filtration (UF) was 3,3 0,4 kg in HD and 3,1 0,2 kg in HDF. Decrease in blood volume HD=10 1% in HD and 11 1% in HDF. Kt/V in HD was 1,6 0,1, and 1,8 0,1 in HDF.
In conclusion, a decrease in SV during HDF in high calcium treatment at temp. controlled, UF and Kt/V matched conditions was found. CO decreased equally in both modalities. MBP did not decrease significantly in any modality. TPR increased only during HDF, and no TPR difference between HD and HDF could be demonstrated.

Sunday, October 31, 2004 , 10:00 AM

Poster: Hemodialysis: Body Composition and Hemodynamics (10:00 AM - 12:00 PM) Poster Board Number: SU-PO275


 

[F-PO399] How Well Does Access Blood Flow (Qa) or Decreases in ABF (Qa) Predict Arteriovenous Graft (AVG) Thrombosis?

Brian M. Murray, Neeraj Singh. Medicine, University at Buffalo, Buffalo, NY

Access surveillance with Qa or Qa has been recommended by DOQI to prolong AVG survival , but it,s utility for this purpose has been challenged (Paulson, AJKD35: 1089, 2000). We conducted a prospective non-interventional study in which 43 patients with functioning AVGs had Qa (by Transonic HD02 meter) measured on 3 occasions, 1 month apart, and were then followed for a further 6 months (without further Qa measurements). Five patients were excluded from the analysis ( 1 AVG clotted before the second measurement, 2 missed Qa measurements due to hospitalization, 2 underwent fistulograms(with angioplasty) for clinical indications). None of the other 38 AVGs had any intervention during the 6 month study period except for AVG thrombosis. There were 7 episodes of thrombosis (18%). We compared the ability of either a Qa < 600ml/min, a Qa >20%, the presence of both criteria or either criterion to predict thrombosis over a 6 month follow-up period. Table 1 shows the sensitivity, false positive rate and post-test probability of thrombosis for each testing criterion.

Predicting thrombosis within 6 mos

Criterion

Sensitivity

False positive rate

PostTest Probability after a PTR

PostTest Probability after a NTR

Qa <600

57%

19%

50%

10%

Qa>20%

86%

13%

59%

3%

Either

86%

29%

39%

4%

Both

50%

9%

55%

18%

PTR=positive test result,NTR=negative test result

It can be seen that Qa performed the best with a sensitivity of 86% and a false positive rate of only 13%, values that would meet those recommended by Paulson. If the presence of either a Qa<600 or Qa >20% were used then there was no improvement in predictive ability because of the increase in false positive rate to 29%. This may have been due to inclusion of patients with low but stable Qa that did not thrombose. On the other hand, requiring that both criteria be satisfied resulted in a decrease in sensitvity (to 50%) with only a minimal improvement in the false positive rate (9%). These data support the proposal of Neyra et al. (KI 54:1714, 1998) that a decrease in Qa may be the best predictor of intermediate term graft thrombosis.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO399

 

 


[PUB195] Clinical Evaluation of Blood Flow Direction in Hemodialysis Graft Protheses.

Roman Fiedler, Matthias Pfau, Bernd Osten, Thomas Langer. Department of Nephrology, Martin-Luther-University Halle-Wittenberg, Halle (Saale), Germany; Department of Internal Medicine, St. Elisabeth and St. Barbara Hospital Halle, Halle (Saale), Germany

The trouble-free use of hemodialysis (HD) shunts is the absolute need for an effective dialysis treatment. Especially in graft loops the confusion of blood flow direction (BFD) with following recirculation can not be excluded in clinical practice, if surgical report is not available. The aim of this study was to prove a simple clinical method for the evaluation of BFD in HD graft protheses.
Forty-seven HD patients, aged 61 [range: 30-80] years, with a graft prothesis were recruited. The mean dialysis duration of the patients was 68 [range: 2-253] months, and the graft prothesis types were Loops (78,7 %), Straights (14,9 %), and Colliers (6,4 %). We investigated for the determination of BFD the simple but safe clinical method of auscultation prior to shunt puncture and compared it to HD01 (Transonic Inc., USA), a conventional ultrasound hemodilution method during HD.
In all grafts blood flow as well as recirculation were measured by HD01, and so BFD was fixed. The shunt noises, which were auscultated by stethoscope (n=47) and recorded as wave file by a microphone computer system (n=32), showed a characteristic stenosis effect by finger press on the graft flow upwards, whereas the noises were scarcely changed by finger press on the graft flow down. Clinical evaluation of BFD was correct in 88 %, impossible in 10 % of the cases and wrong in 1 patient. The study also verified that in 6 patients (12,8 %) dialysis blood tubes were connected always contrary to BFD in the graft. The determined mean recirculation rate was 24,2 %.
From our results we can conclude that the auscultation is a simple sure and easy method for the detection of BFD in grafts, if the investigation is prior to shunt puncture and access blood flow is higher than 400 ml/min. Additional cost and time will be saved in comparison to HD01 and/or duplex sonography.



Publication Only


 

[F-PO400] Why Dont Some Arteriovenous Grafts (AVGs) with Low Access Blood Flow (Qa) Clot?

Brian M. Murray, Neeraj Singh. Dept Of Medicine, University at Buffalo, Buffalo, NY

Previous studies have suggested that a low Qa(<600ml/min) is suggestive of imminent AVG thrombosis, but many AVGs with Qa <600 can maintain patency for a relatively prolonged period of time. The reason for this is not clear. We recently measured Qa and cardiac output (CO) (by Transonic HD02 meter) on 3 occasions, 1 month apart, in 43 patients with functioning AVGs, who were then followed for a further 6 months (without further measurements). Ten of the 43 patients had a value <600ml/min on at least one occasion. Four of these ten AVGs subsequently thrombosed during the follow-up period, whereas the remaining four remained patent. We compared a number of clinical parameters between those AVGs that clotted and those remaining patent (see table 1). The only differences that reached statistical significance were the ratio of Qa to CO and the delta Qa (defined as the % change in Qa between the initial and final measurement).

Table 1

 

Thrombosed AVGs (n=4)

Patent AVGs(n=6)

Patient Age (yrs)

536

6913

Age of AVG (mo)

3339

124

IAP

0.630.29

0.740.21

DVP mmHg

11821

11014

CO L/min

5.40.9

4.61.0

First Qa

687113

533123

Qa/CO

0.080.02

0.12.02*

Delta Qa

-33%17%

-4%14%*

IAP=intra-access pressure ratio, DVP=dynamic venous pressure,*P<.05)

The AVGs that clotted were those that exhibited a fall in Qa during the 2 month measurement period rather than those AVGs that exhibited stable albeit low Qa. As a result, the initial Qa tended to be higher in the AVGs that subsequently clotted.Grafts that remained patent also tended to have a higher Qa/CO ratio, i.e. Qa accounted for a greater percentage of the CO. This may have reflected the fact that the patent AVGs tended to belong to older patients with lower cardiac outputs, though neither of these differences reached statistical significance. It is possible that in these patients the low Qa was in part due to poor cardiac performance as much as a progressive stenosis. In conclusion, while Qa<600 indicates an increased risk of thrombosis; within this group, it is the AVGs with falling Qa that are most likely to clot.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO400

 

 

 


 

 

 


[SA-PO313] Meticulous Medical Care Provides Minimum of Complications and Excellent Survival of Tunneled Central Venous Catheters.

Henrik Mulec, Eva-Lis Henriksson, Elisabeth Fransson, Per Dahlberg. Department of Dialysis and Medicine, Med Clin Northern Alvsborg Hospital, Trollhattan, Sweden

Prospective observational study on complication rate (Infection rate, Blood flow obstruction) and technique survival of tunneled central venous catheters (CVC).
A single center experience.
From 1999 until December 2003 69 patients received 90 ASH split catheters inserted with few exception in right jugular vein with the tunnel to the chest. One operator performed the procedure by ultra sound guidance. Antibiotics were used at the time of insertion. There were 72 primary insertions and 18 catheters were changed in the same vein. There were no per operative complications. Patients were 34 women and 35 men. Mean age was 6514 years and 25% of patients were diabetics. Follow up was 35237 catheter days.
During first 3 months the same nursing team responsible for the patient was responsible for the optimal fixation of the catheter, the team changed the catheter dressing 1/week and after 3 months the dressing was minimal consisting of a tape for fixation. At the start of dialysis the luer lock connection was disinfected by iodopax or chlorhexidin. Heparin 5000 IU/ml was used as locking solution with routine local use of alteplase 1 mg /ml once a month and at blood flow <200 ml/min. Maximal possible blood flow was measured regularly with TRANSONIC technique and was 280 ml/min allowing Kt/V 1.6.
Mean survival time for catheters (dialysis with one and the same catheter) was (Kaplan Meier) 557 62 days. Technique survival, time the patients were dialyzed by CVC was 736 79 days. Catheters not needed and catheters in dead patients were censored. There were few local, exit site infections: 0.7/year or 1/515 catheter days and even less blood stream infections: 0.32/year or 1 /1156 catheter days!
Major long-term complication was the development of vena cava superior syndrome in 3 patients and catheter brake at 4 occasions after 1.5 and 3 years of use respectively.
Good medical care of the tunneled CVC is associated with low infection rate, with excellent catheter survival and blood flows for adequate dialysis.

Saturday, October 30, 2004 , 10:00 AM

Poster: Managing Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: SA-PO313


[F-PO406] Arteriovascular Access Flow Monitoring Via the Ultrasound Dilution Technique Reduces Vascular Access Complications.

Bruno Paris, Sylvie Lavaud, Christine Randoux, Philippe Rieu, Jacques Chanard. Nephrology, University Hospital, Reims, France

Introduction: Hemodialysis vascular access-related complication is an important cause of dialysis patient hospitalization. Sequential prospective monitoring of arteriovenous access flow may detect incipient access malfunction and prevent vascular access morbidity.
Method: We retrospectively studied the incidence of vascular access complications (thrombosis, angioplasty, hospitalisation) during two periods in a chronic hemodialysis population at Reims Hospital. In the first period (2001, 84 patients), access flow was not monitored. In the second period (2003, 142 patients), access flow was assessed every 3 months via the ultrasound dilution technique. Detection of vascular access stenosis was always associated with preventive angioplasty.
Results: During the two study periods: 1) detection of vascular access stenosis was increased from 0,42 (period 1) to 0,53 (period 2) events / patient / year (p< 0,05), 2) vascular access thrombosis rate was reduced from 0,250 (period 1) to 0,126 (period 2) events / patient / year (p< 0,02), and 3) hospital days related to vascular access-related complications was decreased from 1,52 (period 1) to 1,15 (period 2) days / patient / year (p< 0,02). 95% of thrombosis occurred in patients with synthetic arteriovenous fistulas.
Conclusion: Our data shows that detection of vascular stenosis by prospective monitoring of arteriovenous access flow, when coupled with preventive angioplasties, reduces by 50% the risk vascular access thrombosis. Prevention of vascular access morbidity was associated with a decreased in the number of hospitalisations.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO406

 


[F-PO404] Monthly Access Flow Monitoring Did Not Improve Cumulative Access Survival.

Hassan Shahin, Reddy P. Geeta, Bradley S. Franzwa, Bradley S. Dixon. Internal Medicine, University of Iowa College of Medicine, Iowa City, IA; Research, Veterans Administration Medical Center, Iowa City, IA

Regular access monitoring is recommended to detect and treat access stenosis in order to prevent access thrombosis and failure. In 1999, we instituted a program of regular monthly access flow (Qa) monitoring using the ultrasound dilution technique (UDT). In a sequential observational trial, 300 pts were studied for the impact of UDT on survival of the first placed arteriovenous access (native fistulas, AVF prosthetic grafts, AVG). Group 1, the historic group (before 1999), had 204 AV accesses, 72% AVF, followed for 382 access-yrs. Group 2, the UDT monitored group, had 95 AV accesses, 80% AVF, followed for 168 access-yrs. Decision to refer for angiography was based on clinical criteria for Group 1, and clinical criteria plus results of UDT in Group 2; i.e. Qa <400 (AVF) or <600 (AVG) ml/min or Qa>25% in sequential studies. We found that Group 2 had an increased rate of angioplasties (meanSD=1.071.8 vs 0.521.9 per access-yr, p=0.005) but no change in radiological thrombosis events (0.611.8, Group 2 vs. 0.662.3, p=0.94). Surgical procedures were higher in Group 1 (0.893.0, Group 1 vs 0.371.3 per access-yr, p<0.01). Overall, the primary unassisted patency was not statistically different between the two groups but cumulative patency was better in Group 2 even after adjusting for differences in baseline characteristics (p=0.02). However, the improvement in cumulative patency in Group 2 all occurred prior to initiation of access monitoring (median=139 days after access creation). Comparing outcomes in Group 2 patients after initiation of flow monitoring with Group 1 patients whose access survived at least 139 days, primary unassisted patency tended to decrease after initiation of access monitoring in Group 2 (p=0.09) and there was no difference in cumulative patency after flow monitoring. We conclude that UDT monitoring is associated with an increased incidence of angioplasty procedures with a trend towards shorter primary patency, but did not decrease the radiological thrombosis rate or improve cumulative access patency.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO404


[SU-FC077] Forearm Blood Flow Reserve Predicts Successful Maturation of AV-Fistulae.

Joke van der Linden, Anton H. van den Meiracker, Dammis Vroegindeweij, Andre A.E.A. de Smet, Thomas W. Lameris, Marinus A. van den Dorpel. Internal Medicine, Medical Center Rijnmond-Zuid, Rotterdam, Netherlands; Internal Medicine, Erasmus Medical Center, Rotterdam, Netherlands; Radiology, Medical Center Rijnmond-Zuid, Rotterdam, Netherlands; Surgery, Medical Center Rijnmond-Zuid, Rotterdam, Netherlands

The autologous AV-fistula is the vascular access of choice in hemodialysis patients because of superior long-term survival due to a lower risk for thrombosis and infection.
Access surgery should be followed by adequate fistule maturation, i.e. a gradual increase in vessel diameter, and thus increase in access flow. We hypothesised that preoperative measurement of the arterial forearm blood flow (FBF) reserve predicts successful fistula development. Prior to surgery, we measured FBF during incremental infusion of both an endothelium-dependent vasodilator, metacholine (MCh), and an endothelium-independent vasodilator, sodiumnitroprusside (SNP), using venous occlusion plethysmography. Twentyone patients (age 62 3,(mean sem), 10M /11F ) with end-stage renal failure awaiting access surgery were studied. In patients with successful AVF maturation (n=17) the AUC of both vasodilators was higher than in patients with AVF failure (n=4)( MCh: 122.29.8 U vs. 79.15.8, p<0.05; SNP: 255.919.4 vs. 127.829.4, p<0.01). The absolute FBF increase was also significantly different between both groups (MCh:10.30.9 vs. 6.21.0 mL/100mL/min, p=0.03; SNP: 11.81.1 vs. 5.41.4 mL/100mL/min, p<0.01). No AVF failure was observed when AUC was above 100 U after MCh (n=11)(p<0.05), or when AUC was above 200 U after SNP (n=13) (p<0.01). Baseline MAP and FBF was not different (1074 vs. 1095 mmHg, N.S.; 3.40.7 vs.4.60.4, N.S.). Duplex radial artery diameter did not differ between patients with successful or unsuccessful AVF maturation (2.40.2 vs 2.80.3, N.S.). We conclude that with measurement of FBF reserve, successful or unsuccessful AVF maturation can be predicted accurately. Endothelial function does not seem to play a role, as endothelium-dependent and -independent vasodilators had similar effects. In addition to anatomical mapping, functional preoperative evaluation may improve AVF outcome.

Sunday, October 31, 2004 , 4:30 PM

Free Communication: Prevention and Correction of Access Problems (4:00 PM-6:00 PM)


[F-PO530] Extravascular Lung Water and Peripheral Volume Status in Hemodialysis Patients with a History of Heart Failure and without.

G. Joseph, J. M. MacRae, A. P. Heidenheim, R. M. Lindsay. Medicine, University of Western Ontario, London, ON, Canada; University of British Columbia, Vancouver, BC, Canada

Hemodialysis (HD) patients are generally extracellularly volume expanded. HD patients with a history of congestive heart failure (CHF) may be more so. A technique to measure extravascular lung water (EVLW) using blood ultrasound velocity changes following injections of 0.9% and 5% saline has been described. We used this technique in stable HD patients and determined their range of EVLW. We questioned the relationship between volume expansion and EVLW in HD patients with and without a history of CHF and in those with CHF.
We studied 29 stable HD patients, 12 with a history of CHF and 17 without. All were clinically euvolemic. Intracellular fluid (ICF) and extracellular fluid (ECF) were measured prior to initiation of HD using bioimpedance spectroscopy. EVLW was measured in early dialysis using blood ultrasound dilution. The values for EVLW were 3.55 ml/kg 0.94 (M SD) in patients without a history of CHF and 3.88ml/kg 0.82 in patients with a history (NS). These values are similar to those previously obtained. The ECF/ICF ratio was higher in those with a history of CHF (1.27 0.29) compared to those without (1.04 0.04) (p = 0.013). These ECF/ICF ratios are higher than those found in healthy subjects (0.440.74) indicating ECF volume overload in both groups, but more so in those with a history of CHF. There was a positive correlation between EVLW and ECF/ICF ratios (r =0.54; p=0.003) indicating that ECF volume expansion is associated with increased EVLW. EVLW measurements were made in two patients who were clinically in pulmonary edema. Their EVLW were higher than those not in pulmonary edema (7.95 ml/kg and 5.95 ml/kg, p <0.05).
This study shows: 1) that HD patients with a history of CHF are ECF volume overloaded (even when judged euvolemic), to a greater extent than those without; 2) that the degree of ECF expansion is associated with increasing EVLW volumes even without pulmonary edema and; 3) that eventually ECF volume expansion exceeds limits and pulmonary edema occurs. We surmise that these developing technologies will become of clinical value.

Friday, October 29, 2004 , 10:00 AM

Poster: Complications of Hemodialysis I (10:00 AM - 12:00 PM) Poster Board Number: F-PO530


[SU-FC080] A Six Year Study of Risk of Hemodialysis Graft Thrombosis: An Analysis Based upon the Repetitive Longitudinal Nature of Flow Surveillance.

Sunanda J. Ram, Neville R. Dossabhoy, William D. Paulson. Medicine, LSU Health Sciences Center, Shreveport, LA; Medicine, Medical College of Georgia, Augusta, GA

During graft surveillance, a decrease in blood flow (Qa) is widely used to assess risk of thrombosis. To our knowledge, however, no studies have taken into account the repetitive longitudinal nature of Qa measurements. In this 6 year prospective study, we performed 1,842 monthly Qa measurements by ultrasound dilution on 129 patients. We assessed risk of thrombosis associated with monthly determinations of percentage decrease in Qa (Qa) computed over rolling intervals of 3 months. Qa measurements were not criteria for referral for interventional procedures during the study. Risk of thrombosis was determined during the 30 or 90 days after each Qa. Forty-one thromboses occurred within 30 days of 1,036 Qa determinations; 84 thromboses occurred within 90 days of 986 Qa determinations. The figure shows risk of thrombosis in relation to range of Qa values.

The figure demonstrates that risk of thrombosis in the next 30 or 90 days was small unless Qa was 40%. We also found that 50% of all thromboses were preceded by Qa <20%.
Conclusion: This study shows that risk of thrombosis associated with a single Qa determination is small unless Qa is large. Moreover, half of all thromboses are not associated with a significant decrease in Qa. A surveillance protocol that consists of repetitive longitudinal determinations of Qa does not provide an accurate assessment of risk of thrombosis.

Sunday, October 31, 2004 , 5:15 PM

Free Communication: Prevention and Correction of Access Problems (4:00 PM-6:00 PM)


[F-PO391] The Natural History of Autogenous Radio-Cephalic (RC) Wrist Arteriovenous Fistulas (AVF): A Five-year Prospective Observational Study.

Carlo Basile, Giovanni Ruggieri, Luigi Vernaglione, Alessio Montanaro, Rosa Giordano. Division of Nephrology, Miulli General Hospital, Acquaviva delle Fonti, Italy; Division of Nephrology, Martina Franca Hospital, Martina Franca, Italy

Clinical practice guidelines have supported vascular access (VA) surveillance programs on the premise that the natural history of the VA will be altered by radiological or surgical interventions. The primary objective of this prospective observational study was to follow the natural history of autogenous RC wrist AVFs for 5 years without any preemptive interventions.
We enrolled 52 randomly selected adult prevalent hemodialysis (HD) patients. The protocol prescribed avoiding any surgical or interventional radiological procedures until VA failure (AVF thrombosis or a VA not assuring a single-pool Kt/V>1.2). The subjects underwent yearly assessments of VA blood flow rate (Qa) by means of a saline ultrasound dilution method.
All failures of VA were due to AVF thrombosis: it occurred in 9 cases, i.e. a rate of 0.035 AVF thrombosis per patient-year at risk. A ROC curve, evaluating the diagnostic accuracy of baseline VA Qa values in predicting AVF failure, showed an under-the-curve area of 0.820.05 SD (p = 0.01). The value of VA Qa, identified as a predictor of AVF failure, was<700 ml/min with an 88.9% sensitivity and 68.6% specificity. When subdividing AVFs into two groups according to the baseline VA Qa, 2 out of the 9 thromboses occurred among the AVFs that had baseline Qa>700 ml/min (n = 31), whereas 7 occurred among the AVFs that had baseline Qa<700 ml/min (n = 21). The 5-year cumulative actuarial survival was 75.0% and 26.0%, respectively (log-rank test, p=0.006). The 18 AVFs that remained patent at the end of the 5 years maintained a median Qa>900 ml/min at all time points studied.
In conclusion, this study shows a very low rate of AVF thrombosis per patient-year at risk and a high actuarial survival of autogenous RC wrist AVFs, particularly of those having a Qa>700 ml/min. Thus, a VA Qa<700 ml/min appears to be a reliable cut-off point at which to start a closer monitoring of this parameter which may lead to further investigations and possibly interventions relevant to the function of the AVFs.

Friday, October 29, 2004 , 10:00 AM

Poster: Hemodialysis Access Failure (10:00 AM - 12:00 PM) Poster Board Number: F-PO391


[PUB206] One Year Experience of an Endovascular Program Managed by Nephrologists at an Academic Center.

Arif Asif, Donna Merrill, Florin Gadalean, Galo Garces, Gautam Cherla, Patricia Briones, Jan Tawakol, Cristovao Vieira, David Roth. Nephrology and Hypertension, University of Miami, Miami, FL

Hemodialysis (HD) vascular access (VA) management is a critical aspect of the care of the end stage renal disease population. Surgeons and interventional radiologists have traditionally provided this care. Frequently, VA-related procedures do not receive the highest priority by these specialists. Nephrologists are becoming acutely aware of the need for improvements to VA care. To respond to VA-related issues, we initiated a HD VA program. After completing optimal training in endovascular procedures the nephrologist developed a team to assist with various aspects of the procedures. An endovascular procedure room was established within an outpatient dialysis facility. VA surveillance was implemented using ultrasound dilution technique in order to identify patients with VA dysfunction. Procedures are performed on an outpatient basis and VA teaching is routinely provided before and after the actual procedure. Patients are scheduled on their dialysis or non-dialysis day according to their preference, minimizing inconvenience and transportation issues. Thus far; we have successfully completed 430 procedures in a one year period (06-03 to 05-04). The procedures included percutaneous balloon angioplasty (n=172), vein mapping (n=43), salvage of undeveloped fistulae (n=20), vein obliteration (n=10), declotting procedure (n=24), venous stent placement (n=1), tunneled hemodialysis catheter placement (n=46), angioplasty of fibrin sheath (n=16) and tunneled catheter removal procedure (n=48). Our cumulative procedure-related complications included grade I hematoma (n=6), severed micropuncture sheath (n=1) and pneumothorax (n=1). None of these required surgical intervention. This approach has been enthusiastically received by the patients who routinely had suffered procedure-related treatment delays prior to the initiation of our program. We propose that nephrologists should become more involved in the management of VA. In addition, we feel that outpatient HD facilities are a prime location for initiation of such a program.



Publication Only


[PUB209] Long Term Follow-Up of Percutaneous Treatment Vascular Access Dysfunction.

J.A. Herrero, F. Coronel, J.V. Mendez, E. Santos, J.J. Gallego, C. Gamez, S. Cigarran, N. Calvo, A. Barrientos. Nephrology, Hospital Clinico San Carlos, Madrid, Spain; Radiology, Hospital Clinico San Carlos, Madrid, Spain

Vascular access thrombosis is one of the most morbid problems encountered by hemodialysis patients. The purpose of the study was to assess the long term usefulness of a program for the detection and percutaneous treatment of vascular access dysfunction and to compare between the different types of arteriovenous fistulae (AVF). Detection was based on physical examination, blood flow, venous pressure, dialysis efficiency and recirculation measurements as a indicated NKF-DOQI guidelines. Percutaneous treatment consisting of angioplasty (PTA) or PTA plus stent deployment when the residual stenosis was >50%.
From october-90 to march-03 we have performed 559 procedures in 281 AVF (235 patients). The AVF were: 141 radial-cephalic (RC), 52 brachial-cephalic (BC) and 88 polytetrafluoroethylene grafts (PTFE). The most frequently indicators of dysfunction were decreased blood flow rate in RC (65%) and increased venous pressure in BC and PTFE (72 % and 76 % respectively). Technical success rate for PTA was 77 % in RC, 45 % in BC and 47 % in PTFE ( p<0.001 of RC versus BC and PTFE). With stent deployment technical success rate was 98 % in all types of AVF. The primary patency rate of all AVF at 1 and 5 years was 52 % and 28 % respectively, whereas the assisted patency rate (more of 1 procedure/AVF) was 82 % and 49 %. Reinterventions were 2.75 procedures/AVF in BC, 2.22 procedures/AVF in PTFE and 1.57 procedures/AVF in RC (p=0.01 versus BC and PTFE). The period of reintervention was 170 days in BC, 204 days in PTFE and 304 days in RC (p<0.001 versus BC and PTFE). Assisted patency rate at 1 and 5 years was also different when compare the AVF types: RC 88 % and 79 %, BC 83 % and 16 %, PTFE 73 % and 22 % respectively (p=0.003).
We conclude that a surveillance program of vascular access dysfunction with percutaneous treatment as required provide excellent patency rates in radial-cephalic fistulae. The great frequency of stenosis recurrence specially on brachial-cephalic and PTFE fistulae makes it necessary to reinforce the application of detection and early treatment program.



Publication Only


[SU-FC079] Association of Clinic Vascular Access Monitoring with Clinical Outcomes in Hemodialysis Patients: The ESRD Quality (EQUAL) Study.

Laura C. Plantinga, Bernard G. Jaar, Brad Astor, Nancy E. Fink, Joseph A. Eustace, Michael J. Klag, Neil R. Powe. Department of Medicine, Johns Hopkins University, Baltimore, MD

Early identification of access complications may be associated with improved patient outcomes. We examined whether patient outcomes were associated with clinic-reported vascular access monitoring practices in an incident cohort of 363 HD patients treated at 64 clinics, starting in 1995. Analysis was restricted to patients who survived at least 6 months and who had a first permanent vascular access (arteriovenous fistula or graft) by 6 months after the start of dialysis. The outcomes were access intervention, access failure, 6-month Kt/V, and all-cause hospitalization (average follow-up, 1.8 years) and 2-year mortality. Using Poisson regression (to obtain IRRs) and Cox proportional hazards analyses (to obtain RHs), we examined the association of monitoring practices with patient outcomes, adjusting for potential confounders such as age, race, comorbidity, labs at start of dialysis, and access vintage. After adjustment, patients who received monitoring weekly or more often (52%) were more likely to have an access intervention (RH=1.39, 95% CI, 1.05-1.83) than those who received monitoring less than weekly. However, patients treated at clinics that reported performing vascular access monitoring (80% of patients) were 31% less likely to be hospitalized (IRR=0.69, 95% CI, 0.53-0.89) than those who received no monitoring. Vascular access monitoring was not associated with access failure, Kt/V, or mortality. Interestingly, patients treated at clinics that used dynamic venous pressure for monitoring (26%) were more likely to have an access intervention (RH=1.55, 95% CI, 1.24-1.95) and patients treated at clinics that used urea recirculation (UR) for monitoring (33%) were less likely to be hospitalized (IRR=0.56, 95% CI, 0.40-0.79) than patients not monitored with these methods. Although more frequent monitoring is associated with greater chance of access intervention, monitoring is not associated with access failure. Also, these results suggest that vascular access monitoring in HD clinics may be associated with fewer hospitalizations but not with mortality.
Funding Source: NHLBI

Sunday, October 31, 2004 , 5:00 PM

Free Communication: Prevention and Correction of Access Problems (4:00 PM-6:00 PM)


[SU-FC078] Native Arteriovenous Fistula Blood Flow Surveillance: A Randomised Controlled Trial.

Kevan R. Polkinghorne, Kenneth K.P. Lau, Alan Saunder, Robert C. Atkins, Peter G. Kerr. Nephrology, Monash Medical Centre, Melbourne, Victoria, Australia; Radiology, Monash Medical Centre, Melbourne, Victoria, Australia; Surgery, Monash Medical Centre, Melbourne, Victoria, Australia

Clinical practice guidelines recommend that the preferred method of stenosis surveillance for native arteriovenous fistulae (AVF) is the monthly measurement of AVF blood flow (Qa).
However these recommendations are based on observational studies without an adequate control group. Therefore we conducted a randomised, prospective, double blind, controlled trial to assess whether Qa surveillance results in an increased detection of AVF stenosis over and above currently accepted clinical criteria. 137 patients were randomly assigned to receive either continuing AVF surveillance using current clinical criteria (control, usual treatment) or usual treatment plus AVF blood flow surveillance (Qa surveillance group). The primary outcome measure was the detection of a haemodynamically significant AVF stenosis as determined by angiography. 67 and 68 patients were assigned to the control and Qa surveillance groups respectively. Patients in the Qa surveillance group were twice as likely to have a stenosis detected compared to the control group (HR 2.27, 95% C.I. 0.85 5.98, p=0.09) although this did not reach statistical significance. There was also a trend for a significant stenosis to be detected earlier in the Qa surveillance group (p=0.09, log rank test). Qa was significantly lower in AVF with a significant stenosis compared to those which were normal (median Qa 760 versus 1390 ml/min, p=0.006). However, a reliance on the Qa results alone prior to angiography demonstrated at best moderate prediction of stenosis (area under the receiver operating characteristic curve 0.78, 95% C.I. 0.63 - 0.94). This study demonstrates that the addition of AVF Qa monitoring to clinical criteria doubled the detection of clinically significant stenosis. However the reliance on blood flow criteria alone would miss a number of patients with a significant stenosis. Therefore the addition of Qa screening increases the detection of AVF stenosis but should not be the sole method of surveillance.

Sunday, October 31, 2004 , 4:45 PM

Free Communication: Prevention and Correction of Access Problems (4:00 PM-6:00 PM)

 

 

 

 
Start receiving our
Flow Matters™ Newsletter.




 

 

 

 

 

 

Blue_Printer90

Print Page

Catalog90

View Catalog

 

 

 

 

 Surgery | Critical Care | Research | Hemodialysis | OEM | Publications | Customer Service

 

 

 ©2007 Transonic Systems Inc.  All rights Reserved. Privacy Statement